Waking up to cold, dry air and a bright red notification on your ResMed AirSense display reading “Humidifier Fault: Error 008” means the machine has shut down its heated humidification system. The device can still blow air to maintain your prescribed pressure, but it will not warm the water or provide moisture until the fault condition clears.
Quick Answer
Error 008 indicates that the machine’s heater plate has overheated, failed to detect water, or lost connection with its internal thermal sensor. Turn off the machine, let the heater plate cool for 15 minutes, remove the water tub, clean the base and spring contacts, and slide the tub back into place until it clicks firmly shut.
What Error 008 Means in Plain English
The bottom of your ResMed humidifier chamber sits directly on an anodized metal heater plate. Beneath this plate is a temperature sensor (thermistor) that continuously communicates with the device’s main processor.
Error 008 is a hardware interlock triggered when:
- The heater plate temperature spikes too quickly, usually because the water chamber ran dry or is sitting crooked and not absorbing heat from the plate.
- The machine senses an open circuit or mismatch between the power sent to the heating element and the temperature measured by the sensor.
- The water chamber is not inserted far enough to activate the internal mechanical presence switch or transfer heat evenly.
To prevent fire hazards or melting plastic, the firmware shuts down all current to the base plate while allowing the blower motor to keep running so your therapy continues uninterrupted.
Why Water Tank Alignment Matters
On both the AirSense 10 and AirSense 11, the humidifier tub must sit with flat, uniform contact against the internal plate. If the tank is misaligned by even two millimeters:
- An air gap forms between the heating plate and the metal base of the tank.
- The heating plate quickly overheats because there is no water volume to absorb its thermal energy.
- The safety circuit detects this sudden thermal runaway and immediately triggers Error 008 to protect the chassis from heat damage.
INCORRECT (Error 008 Trigger) CORRECT (Full Contact)
┌────────────────────────┐ ┌────────────────────────┐
│ Water Tub Angled │ │ Water Tub Seated │
│ ▲ │ │ │
└───┼────────────────────┘ └────────────────────────┘
Air Gap│ ══════════════════════════
═══════╧═════════════════════ Heater Plate (Even Heat Transfer)
Heater Plate (Overheats Fast)
Step-by-Step Fix to Clear Error 008
Follow this procedure to safely clear the fault and restore heated humidification:
1. Allow the Thermal Plate to Cool
Do not touch the metal base inside the humidifier recess immediately. It can reach high temperatures when Error 008 trips. Unplug the machine from the wall and wait 10 to 15 minutes for the heater plate to return to room temperature.
2. Remove and Inspect the Water Tub
- Slide the water tub out of the side of the machine.
- Inspect the exterior bottom surface of the tub. It must be clean, dry, and completely flat.
- Check for mineral scale, limescale, or crusty white buildup on the metal heat-transfer plate at the bottom of the tub. Mineral deposits act as thermal insulation, trapping heat on the machine’s plate and tripping the error. Clean stubborn scale by soaking the tub in a 50/50 mix of white vinegar and warm distilled water for 20 minutes.
3. Inspect the Internal Cavity and Gaskets
- Look inside the machine’s empty humidifier bay.
- Wipe the circular heating plate with a clean, dry microfiber cloth. Never use abrasive scrub pads or solvents on this surface.
- Check the rear silicone seal where the tub connects to the internal air delivery tube. If this gasket is twisted, pinched, or pushed out of its retaining groove, it will physically prevent the tank from sliding completely into position. For detailed gasket care, see Leaking Water Tank? How to Fix ResMed Chamber Seal Problems.
4. Reseat the Tank with Proper Technique
- Refill the chamber with room-temperature distilled water up to (but not exceeding) the MAX fill line.
- Slide the tub into the side of the machine evenly. Do not push down at an angle.
- Push firmly until you feel the latch engage or hear a distinct, solid click.
- Gently pull outward on the tank handle without releasing the latch. The tank should have virtually no forward-or-backward play.
5. Power Cycle to Reset the Firmware Lock
Once the tub is seated:
- Plug the machine back into the wall outlet.
- Start therapy from the home screen.
- Check the display. The Error 008 message should be gone, and the humidifier icon should show active heating (wavy heat lines).
What if Error 008 Keeps Returning?
If you have cleaned the plate, confirmed the water tub is seated flat, and allowed the unit to cool, but Error 008 reappears within 60 seconds of turning on the machine:
- Test with a Replacement Tub: Over years of heat cycles, the plastic base of the water tank can warp, bowing the metal plate away from the heating surface. Trying a new tub is the quickest way to rule out tank warpage.
- Internal Thermistor Failure: If the error trips even with a brand-new water tank, the temperature sensor under the internal heating plate has burned out or drifted outside its calibrated resistance range. At that point, the entire internal heater assembly must be replaced by an authorized technician. See Water Not Warming Up? Fixing Your CPAP Heated Plate for additional heating element diagnostic checks.